Spezifikation:
Zulässige Höchstgeschwindigkeit.....40000 (U/min);
Bohrer....∮2.35 Handstückbohrer.;
Übersetzungsverhältnis ...... 1:1
2.Mounting Burs
⑴Turn the Bur Lock Ring all the way in the direction of black arrow to open the chuck. Insert a bur into the chuck.(Fig.2)
⑵Return the Bur Lock Ring to the original position until a click sound is heard(Fig.2)
3.Lubrication.
Supply SPRAY after each use and or before autoclaving.
①Push spray nozzle attachment over the SPRAY nozzle until it firmly seats.
②Shake the can 3-4 times to well mix lubricant and propellant.
③Insert the spray nozzle in the rear of the handpiece and spray for approximately 2-3 seconds until the oil comes out of the handpiece head or nose.
4.cleaning and sterilization:
Cleaning:
Wipe clean the handpiece sheath with alcohol-dipped soft tissue or Baumwolle. Never clean the handpiece in boiling water or in chemicals,with wire brushes, or in ultrasonic cleaner.
Sterilization:
Steam autoclave is recommended .Autoclave sterilization required after each patient as noted below:
①Clean the handpiece sheath as described above.
②Lubricate the handpiece with SPRAT(Ref,3 Lubrication).
③Place handpiece into autoclaving pouch and seal it in accordance with instruction on the pouch.
④Autoclavable up to a max. 135 ℃ (275℉)
ex.)Autoclave for 20 min. at 121℃ (250℉).or 15 min.at 132℃(270℉).